Fluxgate Magnetometer
It can measure the total magnetic flux,So as to get the values of flux density and magnetic field strength,The HM6200 fluxmeter is very valuable in magnetizing machine, manual and automatic magnet detection and classification, hysteresis loop measurement system,Hall probe can be connected to test the magnetic field strength more accurately
The third generation HM6200 Fluxgate Magnetometer intelligent high precision fluxmeter/gauss meter is controlled by microprocessor,Using the principle of electronic integration, automatic digital suppression adjustment drift system,An instrument that displays the magnitude of magnetic flux and the intensity of magnetic field measured with a color display screen,
It can measure the total magnetic flux,So as to get the values of flux density and magnetic field strength,TheHM6200 fluxmeter is very valuable in magnetizing machine, manual and automatic magnet detection and classification, hysteresis loop measurement system,Hall probe can be connected to test the magnetic field strength more accurately,
It’s all digital multiplex.Provides digital rs-232, USB and relay signal output,Equipped with various helmholtz coil and single coil.External control switch can be connected to reset..External control switch can be connected to reset. BH CURVE Magnetic Hysteresis Loop Tracer Experiment Apparatus
* reset time can be set automatically reset * digital automatic adjustment drift
Measurement specification
|
measured parameters
|
Coil input |
1 |
Magnetic flux measurement accuracy: |
readable±0.50%±0.05%
range |
Probe input |
1 |
Magnetic intensity measurement accuracy: |
readable±0.50%±0.05% range |
Displays refresh rate |
3Times/SEC. |
Magnetic flux (F) measurement range |
20mWb;200mWb;2000mWb |
Coil compatibility |
HM Coil or homemade coil |
Magnetic intensity (B/M) measurement range: |
0-3.0T(30kG) |
Probe compatibility |
HM Series probe |
Magnetic moment (M) : |
mWb.cm |
The probe function |
Automatic identification and zero correction |
Flux measurement resolution |
1µWb,10µWb,100µWb |
Coil connector |
diameters 3.8mmplug-in |
Resolution of magnetic intensity measurement: |
0.1GS (0.01mT) |
Head connector |
15D (DB15) |
temperature coefficient
|
readable±0.03%/oC |
Communication interface 232 function baud rate:19200,57600,115200 |
input impedance: |
1KΩ,10KΩ,100KΩ,1000KΩ |
Connector: 9-pin type D (DB9), DCE setting (connect to PC directly by cable) optional: remote control system or magnetic field control system (aviation or column connector) |
1:Power interface and switch
2:Industrial control interface
3:RS-232
4:Offset adjustment knob
5:Hall probe interface
6: Industrial control terminal
7: Magnetic flux line through
The HM6200 fluxmeter USES 320*240 color LCD screen.In normal operation, the first line of the monitor displays the currently selected range (2mWb;20 MWB;200 MWB;2000mWb) and measurement mode (F: magnetic flux measurement;B: magnetic field strength;M: magnetization);The second line shows the status prompt for holding and clearing zeros;The third line shows the size of the measured value. The measured value has different meanings and units according to the selected measurement mode, as shown in the following table:
Measurement model
|
Meaning of measured value
|
Optional unit
|
F
|
Magnitude of magnetic flux
|
mWb,KMT
|
B
|
Magnetic field intensity
|
mT,Gs,A/m
|
M
|
Magnetic moment
|
mWb.cm
|